Beverage safety: ensuring pure production water at beverage plant

Tuesday, 16 July, 2024 | Supplied by: Hach Pacific Pty Ltd

Beverage safety: ensuring pure production water at beverage plant

In this case study, a major beverage company wanted to ensure that no excess chlorine went into the reverse osmosis (RO) system and that there was no chlorine in the post-RO water going into production. Source water is from the city tap dechlorinated by GAC before being filtered by RO. Demand-driven production flow and intermittent RO operation presented challenging conditions for the analyser and complicated the data analysis. The target maximum residual chlorine amount in the RO permeate, set by the site, was 40 ppb.

Innovative ULR chlorine measurement techniques

The Hach Ultra-Low Range CL17sc was installed both before and after the RO skid. The post-RO measurement was more critical to product quality and was the focus of the test. Post RO measurements from the ULR CL17sc (colorimetry) were compared to measurements from the Hach DR1300 FL handheld ULR chlorine meter (fluorescence).

Achieving quality control under intermittent flow conditions

Despite the demand-driven intermittent flow conditions, the data demonstrate that the dechlorination and RO processes are under control and excursions above the target chlorine limit (40 ppb) are infrequent and time-limited (1-2 ULR CL17sc measurement cycles). Spikes in chlorine concentration showed after the flow to the RO system was stopped and then restored and are likely caused by water accumulation in the lines and washing residual chlorine out of membranes (Figure 1).

Given the infrequent and short-lived nature of such events (2.5–5 min), the dilution rate of the elevated chlorine concentrations is significant due to combining effluent from all individual membrane cartridges in a pipe and/or holding/equalisation tank prior to bottling, and as confirmed by the average residual. Therefore, the spikes can be ignored when correlated with restart of the RO skid operation.

The shorter the periods of no-flow, the lower the observed spikes, so while it is still not detrimental for the membranes or the production water, this issue can be fully resolved by maintaining as continuous operation of the RO system as possible.

The ULR CL17sc measures total chlorine while the DR1300 FL can measure either free or total chlorine, or bisulfite. The total chlorine measurements between the two instruments were compared (Figure 2) and a free chlorine measurement was also taken on the DR1300 FL to confirm the species present in the finished water.

The free chlorine test on the DR1300 FL confirmed the absence of this species, indicating a prevalence of chloramines in the water after dechlorination with GAC, which agrees with the general knowledge of such an application. The total chlorine verification test confirmed the accuracy of the ULR CL17sc with a more accurate fluorescent reference method.

The ULR CL17sc has LOD of 8 μg/L (ppb) and accuracy of ±5% or 10 ppb, whichever is greater. The CL17sc has LOD of 30 μg/L (ppb) and accuracy of ±5% or 40 μg/L (ppb), whichever is greater. Depending on a user’s targeted chlorine level, specified LOD or accuracy yields, actionable insights can provide a user with confidence that it is below their residual target. Without such confidence tied to the right specifications, users may be breaching their target, putting a system at risk, or incurring additional cost by overfeeding dechlorinating chemicals.

For more information visit https://au.hach.com/beverage.

Online: www.au.hach.com
Phone: 03 9545 6229
Related Products

Tecpro ILMAP filter nozzles

ILMAP has a range of filter nozzles designed to meet the diverse needs of industrial and...

SMC Corporation HF2A-LEY series electric actuator

SMC Corporation has added a rod-type electric actuator to its range — specifically designed...

Edwards Vacuum nEDC 65-150 dry claw vacuum pumps

Edwards Vacuum has added to its nEDC 300 pumps range with the nEDC 65-150 models of dry claw...


  • All content Copyright © 2024 Westwick-Farrow Pty Ltd